omron

VT-X750

In-line X-Ray Inspection System

A high resolution 3D CT scan with the VT-X750 cuts hidden solder joints into hundreds of cross-section slices within a matter of seconds. The ability to isolate individual layers allows the automated inspection algorithms to measure a variety of features without the shadowing or interference of components on the opposite side (including Head-in-Pillow and voiding defects). Omron’s offline repair and programming software also reduces the need to interrupt production and allows users to make adjustments on the fly.

axi 3d x-ray röntgen

Item VT-X750 V3-H VT-X750 V3-C VT-X750-XL V3
Model VT-X750 VT-X750-XL
Type V3-H V3-C V3
Inspection object BGA/CSP, inserted components, SOP, QFP, transistors, R/C chips, bottom-side terminal components, QFN, Power devices, POP, Press-fit CN, etc.
Inspection items Void, open, non-wet, Solder Volume, shifting, foreign object, bridging, Solder fillet, TH Solder filling, Solder ball, etc. (selectable to applications)
Imaging system Method: 3D-slice imaging by using parallel CT
Resolution: 6,8,10,15,20,25,30µm/pixel (selectable in the inspection program) Resolution: 3,6,8,10,15,20,25,30µm/pixel (selectable in the inspection program) Resolution: 10,15,20,25,30µm/pixel (selectable in the inspection program)
X-ray source: Micro-focus closed tube
X-ray detector: Flat panel detector
PCBA Size: 50x50~610x515 mm (2x2 to 24x20 inch) Thickness: 0.4~5.0mm (0.4~3.0mm in 3µm resolution) Size: 100 x 100~1200 x 660 mm Thickness: 0.4~15.0 mm
Weight: Less than 4.0 kg, Less than 10.0 kg (option) Weight: Less than 15 kg
Component clearance: Maximum top: 40 mm, 90 mm (option) Maximum bottom: 40 mm Component clearance: Top: 40 mm, Bottom: 50 mm
Warpage: Less than 2.0 mm (Less than 1.0 mm in 3µm resolution) Warpage: Less than 3.0 mm
Main body Footprint: 1,550(W) x 1,925(D) x 1,645(H) mm Footprint: 2,180(W) x 2,510(D) x 1,735(H) mm
Weight: Approx. 3,100 kg Weight: Approx. 5,350 kg
Conveyor height: 900 ±20 mm
Power supply: Single phase, 200 to 240 VAC, 50/60 Hz
Rated power: 2.4 kVA Rated power: 2.58 kVA
X-ray leakage: Less than 0.5 µSv/h
Air supply: 0.4 to 0.6 MPa
Safety standard: CE, SEMI, NFPA, FDA Safety standard: CE, FDA